Edwards APG100-XM NW16 Active Pirani Vacuum Gauge with Calibration Certificate, D0260100C


Edwards APG100 series Active Pirani vacuum gaugesare available in 2 models. The
APG100-XM is the standard model and measures to 10-3 mbar. The APG100-XLC is a
corrosion resistant version with measurement to 10-4 mbar.

Both Edwards APG100 gauges feature compact size for easy installation, a linear output, and a replaceable
sensor tube. The new gauges are compatible with all Edwards Vacuum TIC instrument
controllers and other active gauge controllers and displays. They are also CSA, C/US
approved as well as fully RoHS compliant due to their lead-free construction

Features and Benefits;

Cable connections and gauge adjustment conveniently located, thereby minimizing the space envelope required for access
Sensor tube can be baked to 150 ºC
Adjustable set-point for simple process control and interlocking
Remote calibration possible
Linear output – 1 volt per decade for easy interface with vacuum control systems
NW16 and NW25 flange options for easy connection to vacuum systems – other flange options on request
LED status indicator shows normal and fault conditions

Login to see price!1

Mass85 g
Internal volume5 cm3
Enclosure ratingIP40
Measurement range (APG100-XM)Atmosphere to 10-3 mbar
Measurement range (APG100-XLC)Atmosphere to 10-4 mbar
Accuracy (APG100-XM)Typically +/- 15% at <100 mbar
Accuracy (APG100-XLC)Typically +/- 15% at <10 mbar
Maximum over-pressure10 bar absolute
Operating temp range5o to 60o C
Storage temp range-30o to 70 oC
Bake-out with no electronics150 oC
Humidity80% RH up to 31 oC decreasing linearly
to 50% RH at 40 oC and above
Maximum altitude3000 m
Filament temperature100 oC above ambient
Electrical supply voltage15 to 30 V d.c. nominal
13.5 V d.c. minimum
32 V d.c maximum
Power consumption1 W
Output signal0 to 10 V d.c. nominal
Set-point – open collector transistor
  Rating30 V d.c. 100 mA
  Range of set-point1.8 to 9.2 V dc
  Fixed hysteresis500 mV (1/2 decade)
  Level setting resolution6 mV