Pressure Range | |
AIM-X and AIM-XL | 10-2 to 10-9 mbar |
Accuracy* | Typically ±30% |
Maximum overpressure | 10 bar absolute (145 psi) |
Power supply | +13.5 to +36V d.c. (max 1 V ripple) |
Power consumption | 2 W Maximum |
Output signal | 2 to 10 V d.c. |
Set point | Open collector transistor |
Maximum voltage | 40V d.c. |
Current | 100 mA max |
Temperature range | |
Operating | +5 to +60°C |
Storage | 0 to +70°C |
Baking (DN40CF version only) | 300 o C with electronics removed DN40CF only |
Materials exposed to vacuum | |
NW25 versions | Stainless steel 304 & 306 & 347, fluoroelastomer, glass |
DN40CF | Stainless steel (AISI 304 and 306), copper, ceramic, very small amount of copper braze |
Internal volume | 26 cm3 |
Weight | 0.81 kg |
AIM-X NW25 | 810 g |
AIM-XL NW25 | 860 g |
AIM-X DN40CF | 1090 g |
AIM-XL DN40CF | 1140 g |
External interface connector | 8-way FCC68/RJ45 socket |
Vacuum fitting | NW25 |
Standards | |
Electromagnetic compatibility | EN 61326 (Class B Emissions) |
Enclosure rating | IP40 |
Edwards AIM XL DN40CF Inverted Magnetron Gauge, D14665000
The Edwards AIM Active Inverted Magnetron Gauges (AIM) combine the gauge-head and controller in one compact Active AIM XL unit. Low Magnetic field version XL for sensitive applications. These gauges have proved to be rugged and reliable in a wide range of applications ranging from scientific instruments to industrial processes.
This gauge has a low external magnetic field making it suitable for use with sensitive analytical instruments or maglev turbo pumps. In addition, this gauge features a linear output for easy integration with a computer or PLC.
Technology
The Edwards AIM-XL-DN40CF Gauge is an inverted magnetron gauge head and gauge controller in a single compact unit. The gauge operates as a cold cathode ionisation gauge, in which pressure is measured indirectly as a function of the current that flows in a Townsend discharge maintained in the body tube.
The measurement range of the AIM-XL-DN40CF Gauge is 1 x 10-9 to 1 x 10-2 mbar.
The Edwards AIM Gauge requires a 13.5 to 36 V d.c. power supply: it has a 2 to 10 V d.c. analogue output which is related to pressure.
The AIM-XL-DN40CF Gauge is compatible with all of Edwards Active Gauge Controllers and with the appropriate versions of the Active Gauge Displays. Alternatively you can use an independent power supply for the AIM Gauge and you can read the AIM Gauge output signal with a voltmeter or an analogue-to-digital converter.
Applications
All vacuum applications requiring rugged and reliable pressure indication in the range 1 x 10-2 to 1 x 10-9 mbar. Typical processes range from general vacuum through industrial coaters and furnaces to scientific instruments and semicon.
Login to see price!1